Ivoclar Programat P100 Manual English -
Measure the resistance of the muffle. If open circuit, replace the QTK muffle. Damaged Pt-Rh platinum thermocouple.
It is widely used in dental laboratories for producing crowns, veneers, inlays, and onlays.
Connect the unit directly to a grounded wall outlet. Do not use extension cords or power strips, as the high wattage can cause voltage drops or overheating.
If it did not melt, or melted completely into a ball before reaching 961°C, use the calibration adjustment screw or menu setting on the unit to offset the temperature display. Routine Care ivoclar programat p100 manual english
Operating the P100 centers around tailoring parameters to match the ceramic manufacturer’s firing charts. How to Program a Firing Cycle
Adjust the time it takes for the furnace head to close.
When the Programat P100 detects a system fault, it displays an error code. Error Code Potential Cause Recommended Action Thermocouple broken or disconnected. Measure the resistance of the muffle
3. Key Components and Features (Programat P100 Manual English)
Visit the Ivoclar Vivadent website and navigate to the “Dental Professionals” or “Lab” section. Register for a free account. Search for “Programat P100” under technical documentation. Often, archived manuals are available in PDF format.
: How to navigate standard settings, toggle between Celsius and Fahrenheit, and adjust buzzer signals. It is widely used in dental laboratories for
To alter parameters, press the specific parameter key (e.g., for Temperature), adjust the value using the arrow keys, and press Enter or Store to save. Place your ceramic object on the firing tray.
Press . The furnace head will smoothly lower, execute the pre-heating, fire according to the designated curve, and lift automatically once the cooling phase begins. 5. Maintenance, Calibration, and Troubleshooting